Validation of Rivaroxaban Assay for US Registration
- Conditions
- Patients Receiving Rivaroxaban Treatment
- Interventions
- Device: STA- Rivaroxaban Calibrator&Control
- Registration Number
- NCT02333929
- Lead Sponsor
- Diagnostica Stago
- Brief Summary
The objective of the study is to demonstrate the performances of the STA® - Rivaroxaban Calibrator and STA® - Rivaroxaban Control in combination with the STA® - Liquid Anti-Xa to determine the quantity of rivaroxaban in plasma samples by measurement of its direct anti-Xa activity.
- Detailed Description
The study is a method comparison between a reference method (LCMS) and the candidate device.
Between 90 and 120 samples obtained from patients treated by rivaroxaban will be collected on 3 sites in the US according to the inclusion and exclusion criteria detailed in the protocol. These samples will be frozen after the collection and sent to the hemostasis laboratories for testing and to a LCMS central laboratory validated for rivaroxaban measurement.
Results expressed in ng/mL obtained with STA® - Liquid Anti-Xa, used in combination with STA® - Rivaroxaban Calibrator and STA® - Rivaroxaban Control will be compared to the result obtained by the LCMS laboratory.
Diagnostica Stago will analyse the results thanks to regressions according to CLSI EP9-A3 recommendations (guideline dedicated to method comparison for quantitative assay) in order to support the equivalence between the 2 methods.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 100
Between 30 and 40 samples/site obtained from patients treated with rivaroxaban.(ideally 10-15 samples per indication and per site depending on the site practices):
- Prophylaxis of VTE in patients undergoing knee or hip replacement surgery (10 mg OD): orthopaedic department of the hospital will be in charge of the sample collection.
- Treatment of deep vein thrombosis (DVT), pulmonary embolism (PE), and risk reduction of DVT and PE recurrence; (15 mg BID for 3 weeks then 20 mg OD): different departments of the hospital will be in charge of the sample collection (samples can come e.g., from emergency room, vascular lab or cancer unit).
- Stroke prevention and reduction of systemic embolism in non-valvular patients with atrial fibrillation (SPAF) (20 mg OD): cardiology department of the hospital will be in charge of the sample collection.
- Patients less than 18 years old
- Patients under other anti-coagulant treatment
- Samples that are not collected, stored, or handled in accordance with sample collection procedures defined in CLSI H21-A5
Study & Design
- Study Type
- OBSERVATIONAL
- Study Design
- Not specified
- Arm && Interventions
Group Intervention Description VTE prevention STA- Rivaroxaban Calibrator&Control Prophylaxis of VTE in patients undergoing knee or hip replacement surgery (10 mg OD): orthopaedic department of the hospital will be in charge of the sample collection. SPAF STA- Rivaroxaban Calibrator&Control Stroke prevention and reduction of systemic embolism in non-valvular patients with atrial fibrillation (SPAF) (20 mg OD): cardiology department of the hospital will be in charge of the sample collection. DVT/PE treatment STA- Rivaroxaban Calibrator&Control Treatment of deep vein thrombosis (DVT), pulmonary embolism (PE), and risk reduction of DVT and PE recurrence; (15 mg BID for 3 weeks then 20 mg OD): different departments of the hospital will be in charge of the sample collection (samples can come e.g., from emergency room, vascular lab or cancer unit).
- Primary Outcome Measures
Name Time Method Number of participants tested once during their rivaroxaban treatment 1 day
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (1)
Lehigh Valley Health Nrwork
🇺🇸Allentown, Pennsylvania, United States