MedPath

Performance Measurement and Safety Evaluation of the PYTHEAS® ODYSSEE Angular Assistance System for Intra-pedicular Screw Placement in Adult Spine Surgery

Not Applicable
Completed
Conditions
Pedicle Screw
Surgical Navigation Systems
Interventions
Device: Guided placement with PYTHEAS® ODYSSEE
Registration Number
NCT05565131
Lead Sponsor
Centre Hospitalier Universitaire de Nīmes
Brief Summary

Pedicle screw malposition is a common problem for spine surgeons. A poorly placed screw can lead to potentially serious complications (hemorrhage, paralysis) in the short- or medium-term. Accurate images during surgery facilitates preoperative planning and aids the placement of pedicle screws.

Surgical navigation covers all techniques and equipment (hardware and software) needed to perform a computer-assisted procedure with 3D image visualization of the surgical act.

Surgical navigation systems have been on the market for about 15 years. Surgical navigation increases the accuracy of pedicle screw placement. Many technologies are available, but they are expensive and often require extensive instrumentation.

The PYTHEAS® ODYSSEE angular assistance system has many advantages over more widely used systems. Indeed, this system provides surgeons with the necessary information to correctly insert pedicle screws during posterior spondylodesis procedures for the treatment of deformity, osteoarthritis (degenerative spine) and trauma.

Compared to competing solutions, the purpose of this device is to reduce pedicle screw malpositioning while offering minimal radiation exposure and the least possible impact on the conventional surgical workflow. It is a lightweight, efficient and easy-to-use medical device with 2-step functionality: 1) Surgical planning of pedicle screws based on a preoperative scan; 2) Angular assistance tool during surgery.

The study investigators wish to evaluate the performance and safety of the PYTHEAS® ODYSSEE medical device (not CE marked) as well as its usability in a pilot study of first use in humans.

Without a navigation system, the literature reports that on average 9 to 10% of perforations are incorrectly positioned by the "free hand" method. With the PYTHEAS® ODYSSEE angular assistance device, the study hypothesis is that this rate will be 5%, i.e. a 50% improvement.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
14
Inclusion Criteria
  • Patient with indication for posterior arthrodesis of at least 2 lumbar or thoracic vertebrae with placement of pedicle screws at vertebral level for :

    • Traumatology: fracture with significant vertebral compression or major kyphosis deformity, ligament and/or disc injury resulting in instability, with the need to free the marrow or nerve roots, OR
    • Deformity: idiopathic scoliosis with surgical indication for increased curvature, thoracic curvature greater than 45° or lumbar curvature greater than 35°, OR
    • Revision surgery for extension of the assembly in the presence of a possible adjacent syndrome responsible for instability and/or central or foraminal stenosis, OR
    • Degenerative: symptomatic spondylolisthesis, degenerative scoliotic deformity, narrow lumbar canal with canal release leading to instability, syndrome adjacent to a previously performed montage.
  • The patient must have given their free and informed consent and signed the consent form

  • The patient must be a member or beneficiary of a health insurance plan

Exclusion Criteria
  • Patients with contraindication to screw placement
  • Patients with contraindication to CT scan
  • Silicon allergy
  • The subject is participating in a category 1 interventional study, or is in a period of exclusion determined by a previous study
  • The patient is unable to express their consent
  • It is impossible to give the subject informed information
  • The patient is under safeguard of justice or state guardianship
  • Pregnant, parturient or breastfeeding

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
PAtients requiring pedicle screw implantationGuided placement with PYTHEAS® ODYSSEE-
Primary Outcome Measures
NameTimeMethod
Accuracy of the PYTHEAS® ODYSSEE angular assistance device for directing the pedicleDay 0

Number of perforations with the correct direction/Total number of perforations performed. The correct direction will be determined using a palpator guide before screw placement, and considered satisfactory when the internal, external, superior and inferior walls of the intra-pedicular path are respected

Secondary Outcome Measures
NameTimeMethod
Adverse eventsDay 0

No adverse events are anticipated directly linked to the device

Operating time according to the calibration mode performed (specific mode/global mode)Day 0

Operating time (minutes): time from incision until closing indicated on the management software.

Surgeon's satisfaction with technical quality of the transfer between the PYTHEAS tablet and the computer for each procedureDay 0

5-level Likert scale + one item with open-ended. The questions asked will be:

* Are you satisfied with the transfer time between the tablet and the computer?

* Are you satisfied with the ease of transfer between the tablet and the computer?

Surgeon's satisfaction with tablet interactionDay 0

5-level Likert scale + suggestions for improvement. The questions asked will be:

* Are you satisfied with the responsiveness of the tablet?

* Are you satisfied with the interactivity of the tablet?

* Are you satisfied with the quality of the images displayed by the tablet?

* Suggestions for improvement.

Correct screw positioning rateDay 0

Number of correctly positioned screws (in trans-pedicular) / Number of screws placed. Correct placement determined by CT scan performed immediately postoperatively

Surgeon's satisfaction with the calibration (specific/global)Day 0

5-level Likert scale + one item with open-ended. The questions asked will be:

* Are you satisfied with the specific calibration?

* Are you satisfied with the overall calibration?

* What type of calibration would you prefer to use in practice? In which cases?

Surgeon's reported overall usability of the PYTHEAS® ODYSSEEDay 0

The System Usability Scale (SUS): a 10-item questionnaire where a score of 75 or more is generally considered "good", and 50 to 75 is considered "fair" or "adequate". A score below 50 indicates major problems in terms of satisfaction.

Trial Locations

Locations (2)

Hôpital Saint Joseph

🇫🇷

Marseille, France

CHU de Nimes

🇫🇷

Nîmes, France

© Copyright 2025. All Rights Reserved by MedPath