The Effect of 10 cm Catheter Mount on Malfunction of Capnometers Under the Conditions of Water
- Conditions
- Cardiopulmonary ResuscitationPulmonary EdemaHemorrhage
- Interventions
- Device: No-DCCMDevice: DCCMOther: Water 5ml in tracheal tubeOther: Water 10ml in tracheal tube
- Registration Number
- NCT02072343
- Lead Sponsor
- Hanyang University
- Brief Summary
Capnometer has been used to assess the success of intubation and the quality of cardiopulmonary resuscitation. However, capnometers can malfunction under the conditions such as pulmonary edema and hemorrhage due to the vulnerability of capnometers for water. Investigators hypothesized that the use of 10cm catheter mount will reduce the malfunction of capnometers under the conditions of water.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 25
- Healthy adults without underlying disease
- Lung and heart disease
- Drinking carbonated beverages
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- CROSSOVER
- Arm && Interventions
Group Intervention Description Mainstream capnometer Water 5ml in tracheal tube - Mainstream capnometer Water 10ml in tracheal tube - Microstream capnograph No-DCCM - Microstream capnograph DCCM - Mainstream capnometer No-DCCM - Mainstream capnometer DCCM - Microstream capnograph Water 5ml in tracheal tube - Microstream capnograph Water 10ml in tracheal tube -
- Primary Outcome Measures
Name Time Method the malfunction rate of capnograph 1 day The malfunction of capnograph is defined that the ETCO2 level is out-of the measured range of each patient under water-free conditions. The calculation formula of malfunction rate is as follows: malfunctioning time of capnograph / total experiment time (2min).
1. Water conditions in tracheal tube
* Water 5ml
* Water 10ml
2. Capnometers
* Mainstream capnometer
* Microstream capnograph
3. Comparison group
* No-DCCM group (proximal capnometer of DCCM)
* DCCM group (distal capnometer of DCCM)
Abbreviation : DCCM, Direct connect catheter mount
- Secondary Outcome Measures
Name Time Method The cumulative survival rate of capnometers under the conditions of water 1 day 1. Kaplan-Meier analysis
1-1. Water conditions in tracheal tube
* Water 5ml
* Water 10ml
1-2. Capnometers
* Mainstream capnometer
* Microstream capnograph
1-3. Comparison group
* No-DCCM group (proximal capnometer of DCCM)
* DCCM group (distal capnometer of DCCM)
Abbreviation : DCCM, Direct connect catheter mountThe consistency of ETCO2 between No-DCCM group and DCCM group under the water-free conditions 1 day 1. Bland-Altman plots under water 0ml in tracheal tube.
1-1. Capnometers
* Mainstream capnometer
* Microstream capnograph
1-3. Comparison group
* No-DCCM group (proximal capnometer of DCCM)
* DCCM group (distal capnometer of DCCM)
2. The intraclass correlation coefficient (ICC) under water 0ml in tracheal tube
2-1. Capnometers
* Mainstream capnometer
* Microstream capnograph
2-2. Comparison group
* No-DCCM group (proximal capnometer of DCCM)
* DCCM group (distal capnometer of DCCM)
Trial Locations
- Locations (1)
Hanyang University
🇰🇷Seoul, Korea, Republic of