NCT07459582招募中不适用
Accuracy of Home Lactate Meter and Accu-chek Glucometer in Patients With Glycogen Storage Disease
试验速览
- 阶段
- 不适用
- 状态
- 招募中
- 入组人数
- 10
- 试验地点
- 1
- 主要终点
- To determine if the lactate Plus meter (both capillary and serum sample) readings are within 20% of lab serum lactate reading 95% of the time, in a population of patients with glycogen storage disease type Ia/Ib/XI
研究概览
简要总结
The goal of this study is to determine if the Lactate Plus meter is accurate compared to lab lactate levels, and to determine if the Accu chek guide glucometer is accurate compared to lab serum glucose levels in patients with Glycogen Storage Disease Types Ia, Ib and XI.
To determine this, patient's will have a one-time planned admission to Connecticut Children's for approximately 8 hours and receive hourly blood draws as well as finger-sticks.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Prospective
入排标准
- 年龄范围
- — 至 60 Years(Child, Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patients with diagnosed/confirmed (by liver biopsy or genetic testing) Glycogen Storage Disease Type Ia/Ib (ICD 10 code: E74.01) or XI (74.09).
- •For pediatric participants: Ability of child's parent/legal guardian to understand and the willingness to sign a written informed consent document
- •For adolescents: assent to participate
- •For adults: Ability to understand and the willingness to sign a written informed consent
排除标准
- •Patients with Glycogen storage disease unspecified
- •Patients not meeting inclusion criteria
- •Patients unable to provide consent
- •Patients who decline to be in the study
结局指标
主要结局
To determine if the lactate Plus meter (both capillary and serum sample) readings are within 20% of lab serum lactate reading 95% of the time, in a population of patients with glycogen storage disease type Ia/Ib/XI
时间窗: Hourly x 8 hours
次要结局
- To determine if the Accu-Chek Guide glucometer (capillary sample) readings are within 20% of lab serum glucose reading 95% of the time, in a population of patients with glycogen storage disease type Ia/Ib/XI(Hourly x 8 hours)
研究者
研究点 (1)
Loading locations...
相似试验
已完成
不适用
The latch score in identification and correction of breastfeeding problems- A Prospective interventional studyCTRI/2025/06/088068Shri BM Patil college and research hospital229
尚未招募
不适用
Comparing two simple blood measures to predict recovery in children with severe infectionCTRI/2025/11/097184UNIVERSITY COLLEGE OF MEDICAL SCIENCES GTB HOSPITAL INSTITUTE130
尚未招募
不适用
Latch score as a tool to promote breastfeedingCTRI/2025/02/079892Dr kusum chandrakar118
尚未招募
不适用
Outcomes of elevated lactate levels in patients admitted to intensive care unitsCTRI/2025/07/090636SRI RAMACHANDRA INSTITUTE OF HIGHER EDUCATION AND RESEARCH100
尚未招募
4 期
Non-invasive measurement of blood glucose using an optical deviceCTRI/2024/02/063268Angeline Kirubha S P300