Validation of Indicor Photoplethysmography (PPG) Valsalva Pulse Response vs Non-invasive Blood Pressure

Completed
Conditions
Heart Failure
Interventions
Device: Indicor
Device: Caretaker
Registration Number
NCT04665011
Lead Sponsor
Vixiar Medical, Inc.
Brief Summary

This study designed is compare the pulse response to the Valsalva maneuver between a photoplethysmography sensor (PPG) and non-invasive blood pressure sensor for the purpose of validating the PPG methodology in deriving a value for the Valsalva pulse response. The aim is to show that the PPG method captures the same phase responses as the blood pressure monitor and the primary outcome is the correlation between pulse amplitude ratios for the pulse at the end of the Valsalva to an average baseline pulse.

Detailed Description

With both a photoplethysmography (PPG; Indicor, Vixiar Medical,. Baltimore, MD) and non-invasive blood pressure (Caretaker, Caretaker Medical, Charlottesville, VA) sensors in place on the index and middle fingers, respectively, each patient will be asked to perform three mild, 10-sec Valsalva maneuvers with at least 45 seconds between. Pulse recordings will be collected from each device to compare phases of the Valsalva pulse response including pulse amplitude changes and baseline movement. The primary outcome is the correlation between pulse amplitude ratios for the pulse at the end of the Valsalva to an average baseline pulse.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
121
Inclusion Criteria
  • Clinically stable adults, including heart failure patients visiting an outpatient clinic
Exclusion Criteria
  • Atrial flutter or atrial fibrillation with an irregular ventricular response
  • Significant atrial or ventricular ectopy
  • History of paradoxical emboli
  • Hypertrophic obstructive cardiomyopathy
  • Known intracardiac shunt
  • Known severe aortic valve stenosis
  • Known severe mitral valve stenosis
  • History of embolic CVA
  • Myocardial infarction within one week of intended Indicor testing
  • Uncontrolled hypertension (systolic BP >160mmHg or diastolic BP>100mmHg)
  • Hypotension (systolic BP <90mmHg)
  • Symptomatic bradycardia
  • Known cholesterol emboli
  • Poor left ventricular function with left ventricular thrombus
  • Unstable angina
  • Significant aortic valvular disease
  • Patients weighing less than 40kg

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Arm && Interventions
GroupInterventionDescription
Clinically stable adults, including heart failure patients visiting an outpatient clinicCaretakerObservational to compare simultaneous pulse tracings from PPG and non-invasive blood pressure monitors in capturing the pulse responses to a Valsalva maneuver.
Clinically stable adults, including heart failure patients visiting an outpatient clinicIndicorObservational to compare simultaneous pulse tracings from PPG and non-invasive blood pressure monitors in capturing the pulse responses to a Valsalva maneuver.
Primary Outcome Measures
NameTimeMethod
pulse amplitude ratio5 minutes

The correlation between PPG-based and non-invasive blood pressure pulse amplitude ratio during a Valsalva maneuver.

Secondary Outcome Measures
NameTimeMethod
Phases of valsalva pulse response5 minuntes

The agreement between PPG-based and non-invasive blood pressure pulse tracings during a Valsalva maneuver.