Skip to main content
Clinical Trials/NCT02089711
NCT02089711
Completed
Not Applicable

Multi-center Study for Normal Database of Optic Nerve Head, Retinal Nerve Fiber Layer, and Macula Parameters With the Heidelberg Spectralis OCT, Study 2

Heidelberg Engineering GmbH1 site in 1 country392 target enrollmentJanuary 2014
ConditionsHealthy

Overview

Phase
Not Applicable
Intervention
Not specified
Conditions
Healthy
Sponsor
Heidelberg Engineering GmbH
Enrollment
392
Locations
1
Primary Endpoint
Normative data for ONH dimensions, RNFL thickness, retinal thickness and thickness of retinal layers at the macula.
Status
Completed
Last Updated
10 years ago

Overview

Brief Summary

This study is being initiated to create a normal database including measures of optic nerve head (ONH), peripapillary retinal nerve fiber layer (RNFL), and macula using the Heidelberg Spectralis OCT device. The range will be determined for each structural parameter across normal eyes in Japanese subjects, and the measurements stratified according to age.

Detailed Description

This multi-center, prospective, observational (non-interventional) study is the second in a series to measure structural parameters of the optic nerve head, the peripapillary retinal nerve fiber layer, and the macula using the Heidelberg Spectralis OCT device. This study is conducted in normal Japanese volunteers. The main goal of the study is to provide the range of these structural parameters in normal eyes. The study will include at least 224 (enrollment target: 249) normal volunteers; each study site will recruit 32 to 40 subjects with an age range of 18 to 90 years and an approximately equal number of females and males. All subjects will undergo Spectralis OCT imaging, biometric and ophthalmoscopic examination, disc photography and visual field testing in one single visit. All examinations performed on the subjects are non-significant risk procedures because the medical devices are cleared for marketing in Europe and in the U.S.

Registry
clinicaltrials.gov
Start Date
January 2014
End Date
December 2014
Last Updated
10 years ago
Study Type
Observational
Sex
All

Investigators

Responsible Party
Sponsor

Eligibility Criteria

Inclusion Criteria

  • Subject is not an employee of the eye clinic.
  • Age ≥18 to
  • Able and willing to undergo the test procedures, give consent, and to follow instructions.
  • Healthy eye without prior intraocular surgery (except cataract surgery) and without clinically significant vitreal, retinal or choroidal diseases, clinically significant diabetic retinopathy (subject may have diabetes), or disease of the optic nerve. Small Drusen are acceptable in older subjects.
  • Japanese decent (self-reported).
  • Negative history of glaucoma.
  • Intraocular pressure ≤21mmHg.
  • Best corrected visual acuity ≥0.
  • Refraction between +6 and -6 diopters and astigmatism ≤ 2 diopters.
  • Axial length ≤ 26.0 mm

Exclusion Criteria

  • Unreliable visual field. The reliability indices should be used as guide as well as the perimetrist's notes.
  • Unusable disc stereo photos.
  • Inability to undergo the tests.
  • Insufficient quality of Spectralis OCT images (this is not determined until after Spectralis OCT examination, and is an unusual circumstance).
  • Minimum requirements are:
  • Retina completely included in image frame,
  • Quality Score ≥ 20 in the stored ART mean images, and
  • For ONH-R scan: Center position error ≤ 100 μm.
  • Note: Inability of a subject to perform one of the two scan patterns or insufficient image quality in one scan pattern does not exclude the subject or eye from the study.

Outcomes

Primary Outcomes

Normative data for ONH dimensions, RNFL thickness, retinal thickness and thickness of retinal layers at the macula.

Time Frame: 12 months

This normal database will be used to determine if an unknown subject has measurement values "within normal limits" or "outside normal limits." It is planned to classify an eye "within normal limits" if the measurement value is greater than or equal to the 5th percentile of the normal distribution, and as "outside normal limits" if the measurement value is smaller than the 1st percentile (or larger than 99th percentile) of the normal distribution.

Study Sites (1)

Loading locations...

Similar Trials