Repeatability, Reproducibility, and Efficacy Testing of Automated Hemodynamic Output by Transthoracic Echocardiography

University of Pennsylvania1 个研究点 分布在 1 个国家目标入组 35 人开始时间: 2021年8月11日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
35
试验地点
1
主要终点
Repeatability (intraobserver variability) of hemodynamic parameters obtained from the Lumify Ultrasound with Hemodynamic Automation

研究概览

简要总结

The purpose of this research is to determine if software can accurately predict certain heart parameters from chest wall ultrasound (non-radiation, non-invasive, imaging modality). An approved ultrasound device (Lumify) will be used to obtain images prior to cardiac surgery.

详细描述

This is a prospective, non-randomized, single visit clinical study. No additional interventions or follow-ups to the subject will be conducted as part of the study. All investigational scanning will be conducted using the as low as reasonably achievable (ALARA) principle to ensure the safety of the subjects. As ultrasound imaging is based on non-ionizing radiation, it is generally considered safe when used by appropriately trained health care providers. The study population will be adults. There are no specific enrollment targets, however it is expected that 120 subjects or fewer will be scanned to obtain the required 40 echocardiographic exams to complete this study. Exceeding the enrollment target of 120 subjects will require an amendment to the protocol. It is estimated this study will last for 12 months, at that time it will be evaluated and may potentially be amended.

This study is considered non-significant risk. There is no specific benefit to individuals participating in the study, however the feedback received is used to improve Philips systems and that can lead to increased diagnostic ability in future products.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者
否

入选标准

  • •Patient willing and capable of providing informed consent and physically capable of performing study related activities
  • •Patient undergoing any of the following types of cardiac surgeries:
  • •Aortic valve repair Aortic valve replacement Mitral valve repair Mitral valve replacement Tricuspid valve repair Tricuspid valve replacement Pulmonic valve repair Pulmonic valve replacement Ascending aorta replacement Coronary Artery Bypass Graft (CABG) surgery

排除标准

  • •Pregnant women
  • •Patients known to be under the jurisdiction of the Department of Corrections
  • •Patients with persistent arrhythmias as confirmed by the principal investigator (MacKay) on a caseby-case basis (e.g. atrial fibrillation, atrial flutter, frequent premature ventricular contractions, frequent premature atrial contractions, etc.)
  • •Patients with a contraindication to Pulmonary Arterial Catheter (PAC) monitoring as confirmed by the principal investigator (MacKay) on a case-by-case basis (e.g. tricuspid or pulmonic valve endocarditis, venous thrombosis present in the superior vena cava, right atrium, or main pulmonary artery)
  • •Patients with a contraindication to TEE monitoring as confirmed by the principal investigator (MacKay) on a case-by-case basis (e.g. previous esophageal surgery, prior gastric bypass surgery, esophageal varices).

结局指标

主要结局

Repeatability (intraobserver variability) of hemodynamic parameters obtained from the Lumify Ultrasound with Hemodynamic Automation

时间窗: 08/01/2021 - 08/31/2022

Repeatability of the hemodynamic parameters (e.g. stroke volume, cardiac output, cardiac index) obtained from the Lumify Ultrasound with Hemodynamic Automation at set time intervals of one minute by a single echocardiographer.

Correlation Testing Lumify Ultrasound with Hemodynamic Automation

时间窗: 08/01/2021 - 08/31/2022

Correlation between the hemodynamic parameters (e.g. stroke volume, cardiac output, cardiac index) obtained from the Lumify Ultrasound with Hemodynamic Automation with same hemodynamic parameters obtained from the reference standard of transesophageal echocardiography and pulmonary arterial catheterization by a single echocardiographer.

Reproducibility (interobserver variability) of hemodynamic parameters obtained from the Lumify Ultrasound with Hemodynamic Automation

时间窗: 08/01/2021 - 08/31/2022

Reproducibility of the hemodynamic parameters (e.g. stroke volume, cardiac output, cardiac index) obtained from the Lumify Ultrasound with Hemodynamic Automation at set time intervals of one minute by a multiple echocardiographers blinded to each other's data.

次要结局

  • Clinical Efficacy of hemodynamic parameters obtained from the Lumify Ultrasound with Hemodynamic Automation(08/01/2022 - 08/31/2023)

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Emily MacKay, DO, MSHP

Assistant Professor

University of Pennsylvania

研究点 (1)

Loading locations...

相似试验

已完成
不适用
Assessment of InfraScanner 2000™ in Detecting Subdural and Epidural HematomasCrushing Skull InjuryHead Injuries, ClosedInjuries, CraniocerebralHead TraumaCraniocerebral InjuriesHead Injuries, MultipleHead Trauma,ClosedHead Trauma InjuryHead Trauma, PenetratingHead Injury MajorMultiple Head InjuryHead Injury, OpenHead InjuriesInjuries, HeadHead Injury, MinorTrauma, Head
NCT03353246Duke University500
已完成
不适用
Detecting Traumatic Intracranial Hemorrhage With the InfraScanner 2500™ in UgandaCrushing Skull InjuryHead Injury MajorInjury, HeadHead InjuryHead Injuries, MultipleHead Trauma, PenetratingCraniocerebral InjuryHead TraumaHead Trauma,ClosedTrauma, HeadHead Injuries, ClosedHead Injury, MinorMultiple Head InjuryHead Trauma InjuryHead Injury, OpenInjury, Craniocerebral
NCT06491173Duke University180
已完成
不适用
Assessment of InfraScanner 2000™ in Detecting Subdural and Epidural HematomasHead Injuries, MultipleMultiple Head InjuryCrushing Skull InjuryHead Trauma,ClosedHead Trauma InjuryHead TraumaHead Injuries, ClosedHead Injury, OpenCraniocerebral InjuriesHead InjuriesHead Trauma, PenetratingHead Injury, MinorHead Injury MajorInjuries, CraniocerebralInjuries, HeadTrauma, Head
NCT04505293Duke University672
已完成
不适用
A Study to Evaluate Accuracy and Validity of the "Chang Gung" Ventricular Systolic Dysfunction Screening SoftwareLeft Ventricular Systolic Dysfunction
NCT05861115Chang Gung Memorial Hospital1,172
进行中(未招募)
不适用
A Study of Wearable Devices to Predict MigrainesMigraine
NCT05755945Mayo Clinic20