Moni-Patch Temperature Monitoring Patch System Clinical Accuracy Validation
试验速览
- 阶段
- 不适用
- 状态
- 尚未招募
- 发起方
- 入组人数
- 105
- 试验地点
- 1
- 主要终点
- Agreement between Moni-Patch™ Core Temperature Monitoring Patch System and Continuous Tympanic Temperature (Ttym) Measurements
研究概览
简要总结
Moni-Patch Temperature Monitoring Patch System, a non-invasive, continuous temperature monitoring device designed to estimate core body temperature by applying a Sensor to the neck.
This clinical accuracy validation study aims to verify the accuracy and repeatability of the Moni-Patch compared with an FDA-cleared continuous tympanic temperature monitoring device used as the clinical reference.
Main objectives of this study are; To verify that the Moni-Patch and the reference device(tympanic) demonstrate acceptable agreement, as defined by Bland-Altman Analyses (bias within ±0.4°C and limits of agreement between -1.0°C and 1.0°C).
To verify the consistency and repeatability of temperature measurements obtained from the Moni-Patch during continuous monitoring.
详细描述
This study is conducted in accordance with ISO 80601-2-56:2017, which outlines definitions for temperature-measuring devices, age groups, required sample sizes, fever definitions, procedural flow, and statistical methods for clinical performance evaluation.
All subjects undergo temperature measurement using the reference device(Tympanic) and the investigational device (Moni-Patch Temperature Monitoring Patch System). The Moni-Patch Sensor is applied to the neck and paired with the Receiver, while the tympanic probe is positioned and sealed in the ear canal.
Each subject participates in a continuous 15-minute temperature monitoring session, during which both devices collect temperature values concurrently every 30 seconds.
Environmental temperature and humidity are also recorded during the session.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 7 Years 至 —(Child, Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Age ≥ 7 years
- •Agree to Sensor placement on the neck and ear
排除标准
- •Ear conditions contraindicating tympanic probe; skin issues at placement sites
- •Antipyretic use within 120 minutes; inability to position neck Sensor over carotid
- •Inability to seal tympanic sensor; tympanic membrane not visually confirmed
研究组 & 干预措施
Concurrent Monitoring
Subjects wear Moni-Patch on the neck and tympanic sensor in the ear for concurrent continuous temperature monitoring for 15 minutes after stabilization.
干预措施: Moni-Patch Temperature Monitoring Patch System (Device)
结局指标
主要结局
Agreement between Moni-Patch™ Core Temperature Monitoring Patch System and Continuous Tympanic Temperature (Ttym) Measurements
时间窗: 15-minute continuous temperature monitoring period following a minimum 30-minute sensor acclimation period (per subject).
Agreement will be quantified using the mean temperature difference (bias; °C) and Limits of Agreement (LOA; °C) between the Moni-Patch™ Temperature Monitoring Patch System and the comparative FDA-cleared continuous tympanic thermometer (Ttym). Paired temperature measurements (Moni-Patch vs. Ttym) collected every 30 seconds during continuous monitoring will be analyzed using Bland-Altman methodology. Name of Measurement: Core body temperature difference Measurement Tool(s): Moni-Patc Temperature Monitoring Patch System FDA-cleared continuous tympanic temperature monitoring device (Ttym) Unit of Measure: Degrees Celsius (°C) Analysis Method: Bland-Altman bias and 95% limits of agreement
次要结局
未报告次要终点