Bedside Respiratory Patient Monitoring System - Sensor Off Study

Completed
Conditions
Healthy Volunteers
Registration Number
NCT02075346
Lead Sponsor
Medtronic - MITG
Brief Summary

Use of an investigational oximetry system to evaluate the triggering of the "sensor off" status with the marketed off-the-shelf sensors.

Detailed Description

The Multi-functional Patient Monitoring PCBA is an electronic circuit board in the monitoring system that performs calculations.

The study will evaluate the performance of the Sensor Off New Patient Monitoring system by exposing the sensor to an ambient light source, sensor slipping off the finger with the light emitter and detector facing each other or a sensor slipping off under a blanket.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
18
Inclusion Criteria
  • Male or female subjects 18 years or older (inclusive)
  • Subject is able to participate for the duration of the study
  • Subject is willing to sign an informed consent
Exclusion Criteria
  • Previous injury or trauma to fingers or hands that may change blood flow or vascular supply and affect our ability to test sensors.
  • Physiologic abnormalities that prevent proper application of pulse oximetry sensor.
  • Severe contact allergies that cause a reaction to standard adhesive materials such as those found in medical sensors and electrodes

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
To validate that the device posts a "Sensor Off" message within one minute after the sensor is removed at least ninety percent of the time.The planned duration of the study is 1 month.
Secondary Outcome Measures
NameTimeMethod