Pulse Oximetry- Evaluating Sensor Off- PCBA-1
Completed
- Conditions
- Healthy Subjects
- Registration Number
- NCT01791569
- Lead Sponsor
- Medtronic - MITG
- Brief Summary
To demonstrate that the sensor-off feature on the Pulse oximeter displays per specifications when the sensor is removed from the finger.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 20
Inclusion Criteria
- Male or female subjects 18 or older (inclusive).
- Subject is able to participate for the duration of the study
- Subject is willing to sign an informed consent
Exclusion Criteria
- Previous injury or trauma to fingers or hands that may change blood flow or vascular supply and affect our ability to test sensors
- Physiologic abnormalities that prevent proper application of a medical sensor and electrode
- Severe contact allergies that cause a reaction to standard adhesive materials such as those found in medical sensors and electrodes
Study & Design
- Study Type
- OBSERVATIONAL
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method Device posts Sensor Off within 1 minute after sensor is removed at least 90% of the time. 1-2 hours per subject
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (1)
Covidien