Effect of Cardiac Resynchronization Therapy (CRT) on Left Ventricular Assist Device (LVAD) Function

Not Applicable
Terminated
Conditions
LVAD (Left Ventricular Assist Device) Driveline Infection
Interventions
Other: Temporary changes in an already-implanted medical device (biventricular pacemaker)
Registration Number
NCT04153786
Lead Sponsor
University of California, Davis
Brief Summary

Certain patients with congestive heart failure are treated both by implantation of an LVAD, as well as implantation with a biventricular pacemaker. Both of these devices, individually, have been shown to improve the health of patients with heart failure. However, only a small number of patients have both an LVAD and a biventricular pacemaker at the same time.

Pacemakers have many different settings. However, there is little data to inform physicians which of these different settings is best for the flow of an LVAD. This study is evaluating patients who have both a biventricular pacemaker and an LVAD. Investigators will operate the pacemaker at multiple different settings and monitor the LVAD's function to determine which, if any, setting is best for the LVAD's flow.

Detailed Description

Cardiac resynchronization therapy (CRT) with a biventricular pacemaker has been well-validated to improve symptoms and reduce death in select patients with heart failure. In healthy adults, the two chambers of the heart which pump blood throughout the body (the left and right ventricles) activate and contract together in synchronous fashion. In certain patients with heart failure (specifically, those with a reduced ejection fraction), the left and right ventricles no longer activate and contract synchronously. CRT restores ventricular synchrony by simultaneously activating the left (LV) and right ventricles (RV) with electrical pacing.

Left ventricular assist devices (LVADs) are used to provide mechanical support to patients with more advanced, often end-stage, heart failure. In these cases, the heart's ventricles function so poorly that even despite all standard therapies (including CRT), patients experience symptoms of congestive heart failure even at rest. The LVAD suctions blood from the ventricle and propels it into the aorta, offloading the ventricle and assisting its function.

Due to the similar reasons for using each therapy, a number of patients already possess a biventricular pacemaker at the time of LVAD implantation. Individually, each intervention is known to improve both patient survival and functional status. However, there is sparse data to evaluate the effectiveness of using both devices simultaneously. Specifically, no data is yet available assessing the effect of CRT on the functional parameters of the LVAD.

This protocol will operate the biventricular pacemaker at various settings and assess for changes in the LVAD's function in response to those settings. "Fine-tuning" and identifying the ideal the pacemaker settings will maximize clinical benefit in patients with both devices.

Recruitment & Eligibility

Status
TERMINATED
Sex
All
Target Recruitment
9
Inclusion Criteria
  • All patients who have undergone implantation of both a biventricular pacemaker and LVAD who are at least three months post-LVAD implantation and who receive routine follow-up care at the UC Davis / UC San Diego Adult Arrhythmia/ Device/ LVAD clinic.
  • Potential enrollees have been directly identified by their existing medical providers in clinic, who are investigators in this study.
Exclusion Criteria
  • Any patients who do not present to their routine follow-up appointment at the time of this study.
  • Any patients who have undergone LVAD placement less than three months prior to enrollment.

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
CohortTemporary changes in an already-implanted medical device (biventricular pacemaker)Pacemakers will be programmed to BiV pacing, left ventricular (LV) pacing, right ventricular (RV) pacing and no pacing for five minutes each. LVAD flow will be recorded every thirty seconds for five minutes with each setting. Once the interventions have been completed, all pacemakers will be returned to their original setting.
Primary Outcome Measures
NameTimeMethod
LVAD pulsatility index (PI)2-5 minutes

LVAD PI reflects the contribution of the native heart's cardiac output (or "pulse") on pump flow

LVAD speed2-5 minutes

LVAD speed, or revolutions per minute (RPMs) of the device, is set by providers to a fixed value.

LVAD power2-5 minutes

Power is the voltage required by the device over each cardiac cycle.

LVAD flow2-5 minutes

Flow is the liters per minute (L/min) of blood moved by the device.

Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (1)

University of California Davis Medical Center

🇺🇸

Sacramento, California, United States