Direct Comparison of 2D Cardiac PET With 3D Cardiac PET Using a 3D Monte-Carlo Scatter COrrection Algorithm
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 19
- 试验地点
- 1
- 主要终点
- Quantitative Accuracy
研究概览
简要总结
Lowering the per-infusion dose of Rb-82 offers advantages of lessening radiation exposure and extending useable generator life. Prior studies have not shown equivalence of 3D vs 2D Rb-82 PET. The investigators therefore compare 3D PET after a lower Rb-82 dose (~20 mCi) processed using a Monte-Carlo driven scatter correction algorithm against conventional higher dosage (~50 mCi) 2D Rb-82 PET MPI.
详细描述
This study re-examines 2D an 3D myocardial perfusion PET data from 19 subjects, mean age 65 years, mean BMI 30 kg/m2, 79% male. Rest and stress scans were acquired on a Siemens Accel™ PET scanner in 2D (septa extended) and 3D (septa retracted) modes. Rb-82 doses were rest 53+/-5 mCi and stress 53+/-6 mCi. Imaging times were 2D emission scan for 3 mins (90 sec delay post Rb-82 infusion), followed by a 3 minute, 3D gated emission scan (180 sec delay). Decay and shorter acquisition times led to a 62% reduction in the effective dosage 2D vs 3D datasets. 3D images were first pre-processed using a Monte-Carlo scatter and prompt gamma correction algorithm (Imagen3D™) then reconstructed using ImagenProTM (CVIT, Kansas City, MO). Reconstructed images were evaluated using relative, 17 segment raw scores (Cedars QPET). Studies were read by consensus of 2-blinded readers for: image quality (1-4, poor-excellent), interpretive certainty (1-3, low-high) and rest perfusion using a 17 segment model (0=normal; 1-3 = mild, moderate, or severe perfusion defects). Stress segmental scores were not evaluated due to differences in imaging start time post dipyridamole infusion.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Control
- 时间视角
- Retrospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •>18 Years of Age
- •Have ability to provide informed consent
- •Have a technically adequate rest/stress Rb-82 perfsion PET study within the past 60 days
排除标准
- •Pregnant or nursing feamle
- •Weight > 275 lbs
- •Have contraindications to adenosine stress testing
- •Unstable cardiac rhythm (atrial fibrillation, frequent PVCs
- •PTCA or CABG within the last 60 days
结局指标
主要结局
Quantitative Accuracy
时间窗: 1 day
To establish the quantitative accuracy of reconstructed volumes using Imagen3DTM when compared to other methods.
Visual Assessment
时间窗: 1 day
To establish that visual assessments of image quality and reader confidence using 3D data are equivalent to the other methods.