Accuracy of Pulse Oximeters With Profound Hypoxia NIHO 14

Nihon Kohden1 个研究点 分布在 1 个国家目标入组 12 人开始时间: 2022年3月23日最近更新:
适应症

试验速览

阶段
不适用
状态
进行中(未招募)
发起方
Nihon Kohden
入组人数
12
试验地点
1
主要终点
Root-mean-square (Rms) of device SpO2 measurement with room air stabilization is within +-2% [%SpO2] of reference co-oximeter SaO2 measurement at varying desaturation levels (70% to 100%)

研究概览

简要总结

The aim of this study is to determine the accuracy of devices called pulse oximeters, which measure blood oxygen by shining light through fingers, ears or other skin, without requiring blood sampling. Study will be used with patients at rest

详细描述

This study is intended to evaluate performance of a new design pulse oximeter (test device) manufactured by Nihon Kohden Corporation sufficiently to support performance claims for an FDA 510K submission or ISO technical file. Specifically, SpO2 and pulse rate accuracy will be assessed for a Nihon Kohden OLV-4202 pulse oximeter with adult patients under a controlled setting of varying levels of inhaled oxygen concentration levels for patients at rest.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Prospective

入排标准

年龄范围
18 Years 至 50 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Both male and female subjects who can give written informed consent
  • Healthy subjects capable of undergoing controlled hypoxemia to the levels outlined in the desaturation profile
  • Meeting the demographic requirements

排除标准

  • Pregnant women
  • Significant arrhythmia
  • Blood pressure above 150 systolic or 90 diastolic
  • Carboxyhemoglobin levels over 3%
  • Subjects whom the investigator consider ineligible for the study

结局指标

主要结局

Root-mean-square (Rms) of device SpO2 measurement with room air stabilization is within +-2% [%SpO2] of reference co-oximeter SaO2 measurement at varying desaturation levels (70% to 100%)

时间窗: 30 minutes

Each subject will have an arterial catheter inserted in the left or right wrist and the pulse oximeter study device placed on different fingers of the left or right hand, or the left or right earlobe or the forehead. Measurement position is randomized across subjects. For each subject, a series of desaturation runs are performed. Run 1 stabilized room air period -\> stabilized plateaus at saturation level targets: room air, 92%, 87%, 82%, 77%, 72% Run 2 stabilized 100% O2 period -\> stabilized plateaus at saturation level targets: 100%, 93%, 88%, 83%, 78%, 73% SpO2 is observed breath by breath arterial saturation, which is in turn computed from end tidal PO2 and PCO2. FiO2 is reduced suddenly at first then adjusted to achieve a stable plateau value at the desired target. Each desaturation plateaus should not exceed 10 minutes. A blood draw is taken at stable FiO2 and +30 seconds. The gas mixture is then changed for the next value and repeated for each plateau value.

Root-mean-square (Rms) of device Pulse Rate measurement with room air stabilization is within +-3 [1/min] of reference ECG Pulse Rate measurement at heart rates (30 to 300 [1/min])

时间窗: 30 minutes

Each subject will have an arterial catheter inserted in the left or right wrist and the pulse oximeter study device placed on different fingers of the left or right hand, or the left or right earlobe or the forehead. Measurement position is randomized across subjects. For each subject, a series of desaturation runs are performed. Run 1 stabilized room air period -\> stabilized plateaus at saturation level targets: room air, 92%, 87%, 82%, 77%, 72% Run 2 stabilized 100% O2 period -\> stabilized plateaus at saturation level targets: 100%, 93%, 88%, 83%, 78%, 73% Pulse rate is recorded from the test device and from the reference ECG monitor. PR data measurements are gathered during the stable plateau value at the desired saturation target. Each desaturation plateaus should not exceed 10 minutes. The gas mixture is then changed for the next value and repeated for each plateau value.

次要结局

  • Root-mean-square (Rms) of device SpO2 measurement with room air stabilization is within +-2% [%SpO2] of reference co-oximeter SaO2 measurement at desaturation levels (80% to 100%) and within +-3% [%SpO2] at desaturation levels (70% to 80%)(30 minutes)

研究者

发起方
Nihon Kohden
申办方类型
Industry
责任方
Sponsor

研究点 (1)

Loading locations...