Study to Examine the Correlation of 'Estimated Continuous Cardiac Output' (esCCO) With Transesophageal Echocardiography (TEE) in Pediatric Cardiac Surgery
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 44
- 试验地点
- 1
- 主要终点
- Cardiac Output
研究概览
简要总结
The purpose of this study is to determine the measuring accuracy of the truly non-invasive method estimated continuous cardiac output (esCCO) against the reference method transesophageal echocardiography in pediatric cardiac surgery. esCCO is calculated using the pulse wave transit time, which is defined as the time between the R-wave of the ECG and the 30% increase of the differentiated pulseoximetry waveform.
详细描述
The objective is to compare two methods of cardiac output measurement in children: the new non-invasive method called esCCO™ (by Nihon Kohden, Tokyo, Japan) and the established more invasive method transesophageal echocardiography (TEE). esCCO™ is a completely non-invasive method that could provide a cardiac output measurement in a large range of pediatric patients because of an extremely low risk benefit balance.
Cardiac output monitoring has been technically challenging in pediatric patients. Available methods for measuring cardiac output in children are either inaccurate or highly invasive or limited to a small number of patients. So far esCCO™ has not been tested in the pediatric population but showed promising results when it was tested against the pulmonary artery catheter in adult patients undergoing heart surgery in previous Japanese studies.
Thus the validity of esCCO™ is now tested in pediatric patients scheduled for elective cardiac surgery. Transesophageal echocardiography has been chosen as reference method because it is performed routinely during pediatric cardiac surgery.
The algorithm for esCCO™ -measurement is integrated in an approved patient monitor that replaces the standard patient monitor during the measurement-period.
In the study protocol, events like esCCO™ -calibration time, catheter positioning time, ventilation settings, volume in- and output and drug administration, patients's data and diagnoses are logged. Vital parameters from the patient monitor including the esCCO™-cardiac output values are recorded every second in real time on a study laptop mounted on the same rack as the patient monitor but electrically separated from the other devices.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Prospective
入排标准
- 年龄范围
- — 至 17 Years(Child)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •written consent
- •children <=18 years
- •elective cardiac surgery
- •Sinus rhythm
排除标准
- •no written consent
- •continuous severe cardiac arrhythmias
- •cardiac Pacemaker
- •intraaortal balloon pump
- •contraindications for TEE
- •persistent ductus arteriosus botalli
结局指标
主要结局
Cardiac Output
时间窗: Intraoperative
EsCCO™ measurements shall be compared to TEE cardiac output measurements at three given times: (1) before surgical intervention, (2) shortly after cardiopulmonary bypass and (3) after closure of the thorax
次要结局
未报告次要终点
研究者
Christoph Schramm, M.D.
Principal Investigator
Heidelberg University