Comparison, Calibration, and Validation of the Non-invasive Cloud DX Pulsewave Health MonitorTM With Direct Invasive Radial and Central Arterial Measurements During Cardiac Catheterization
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 20
- 试验地点
- 1
- 主要终点
- Systolic Pressure
研究概览
简要总结
The objective of the Cloud DX Pulsewave Health Monitor study is to prospectively calibrate and validate a novel, non-invasive wrist cuff blood pressure device against the gold standard for hemodynamic monitoring (intra-arterial pressure) in voluntarily consented patient participants who are scheduled for an elective cardiac catheterization procedure for clinically valid reasons.
详细描述
The Cloud DX Pulsewave Health Monitor is a non-invasive wrist cuff blood pressure device that is licensed by Health Canada with Food and Drug Administration approval. Systolic and diastolic pressures are measured via the arterial pulse waveform of the left radial artery. Initially, the blood pressure algorithm was created and calibrated using the auscultatory method. The purpose of this study is to calibrate the device to the clinical gold standard of peripheral and central intra-arterial pressures.
The device acquires a pulse signal in real-time, stores signal and measurement data, and displays stored signal and measurement data for interpretation via internet servers where it is accessible by end-users (physicians and/or patients). The system provides secure accounts for both in-patient and out-patient monitoring. Cloud DX's Pulsewave Health Monitor servers are housed in a secure data center. Privacy measures are to SSAE16 SOC-1 Type-II compliance: 1) restricted secure access, 2) dedicated firewall, 3) HTTPS channel communication via website, 4) self-encrypting hard-drives rendered useless if removed from the server.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Prospective
入排标准
- 年龄范围
- 19 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •≥ 19 years of age
- •Patients recommended by their cardiologist for a first-time, elective cardiac catheterization procedure
- •Wrist circumference between 13.5cm - 23cm (5.3 - 9.1in.)
- •Bilateral arm blood pressure equality (systolic within ±10mmHg; diastolic within ±5mmHg) on 2 separate readings (Nurse Associate & TRA)
- •Willing to volunteer to participate and to sign the study specific informed consent form
排除标准
- •No history of peripheral vascular disease, no previous percutaneous coronary intervention, nor previous coronary artery bypass graft
- •No arrhythmia
- •No abdominal aortic aneurysm
- •No hand/body tremor
结局指标
主要结局
Systolic Pressure
时间窗: Non-invasive blood pressure device measurements on day of cardiac catheterization procedure.
Peripheral arterial pressure (mm Hg)
Diastolic Pressure
时间窗: Non-invasive blood pressure device measurements on day of cardiac catheterization procedure.
Peripheral arterial pressure (mm Hg)
次要结局
未报告次要终点
研究者
Sohrab Lutchmedial, MD, FRCPC
Co-Director of CVR-NB/ Director of NBHC Catheterization Laboratory
Cardiovascular Research New Brunswick