跳至主要内容
临床试验/NCT03124836
NCT03124836已完成不适用

Accuracy of Methemoglobin Measurement of Various Rainbow Pulse Oximeter Sensors

Masimo Corporation2 个研究点 分布在 1 个国家目标入组 28 人开始时间: 2009年11月18日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
28
试验地点
2
主要终点
Accuracy of Methemoglobin Measurement by Arms Calculation

研究概览

简要总结

In this study, the level of methemoglobin (HbMet) will be increased in a controlled manner by administering sodium nitrite intravenously in healthy volunteers. The accuracy of the noninvasive HbMet measurement by a Pulse Oximeter Sensor will be assessed by comparison to blood measurements from a laboratory analyzer.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
18 Years 至 50 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • The subject is male or female, aged ≥18 and <
  • The subject is in good general health with no evidence of any medical problems.
  • The subject is fluent in both written and spoken English.
  • The subject has provided informed consent and is willing to comply with the study procedures.

排除标准

  • The subject is obese (BMI>30).
  • The subject has a known history of heart disease, lung disease, kidney or liver disease.
  • Diagnosis of asthma, sleep apnea, or use of CPAP.
  • Subject has diabetes.
  • Subject has a clotting disorder.
  • The subject has a hemoglobinopathy or history of anemia, per subject report or the first blood sample, that in the opinion of the investigator, would make them unsuitable for study participation.
  • The subject has any other serious systemic illness.
  • The subject is a current smoker.
  • Any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigators' would interfere with the sensors working correctly.
  • The subject has a history of fainting or vasovagal response.
  • The subject has a history of sensitivity to local anesthesia or allergy to the study drugs.
  • The subject has a diagnosis of Raynaud's disease.
  • The subject has unacceptable collateral circulation based on exam by the investigator (Allen's test).
  • The subject is pregnant, lactating or trying to get pregnant.
  • The subject has a history of migraine headaches.
  • Th subject is currently taking antidepressants.
  • The subject is unable or unwilling to provide informed consent, or is unable or unwilling to comply with study procedures.
  • The subject has any other condition, which in the opinion of the investigators' would make them unsuitable for the study.

结局指标

主要结局

Accuracy of Methemoglobin Measurement by Arms Calculation

时间窗: 1-5 hours

Accuracy willl be determined by comparing the noninvasive blood methemoglobin measurement (expressed as a percentage of total hemoglobin) of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.In order to obtain the Arms value, the blood sample methemoglobin value is subtracted from the pulse oximeter methemoglobin value for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor

研究点 (2)

Loading locations...

相似试验