Verify Wrist BPM TMB-1117 to Comply With ANSI/AAMI SP10

Leo Wang1 个研究点 分布在 1 个国家目标入组 85 人开始时间: 2012年5月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
85
试验地点
1
主要终点
Verify the accuracy of measure functions of device

研究概览

简要总结

The clinical protocol of the clinical testing of this device:

  1. Objective of the test: To verify the efficiency of device.
  2. Test methods and procedures: Clinical test in two positions: Seated and Supine.
  3. DUT: Transtek Wrist Blood Pressure Monitor, Model: TMB-1117. Cuff size: 13.5 - 21.5 cm.
  4. Comparison device: Yuyue Medical Blood Pressure Meter, YYBP-212, accuracy: ±1 mmHg and range: 0-300 mmHg.
  5. Study endpoints: Comply with ANSI/AAMI SP10-2002 standard.
  6. Statistical methodology used: Description of statistical methods.
  7. Result: Meet the requirements of SP10.

详细描述

Hospital Information

Hospital Information The data was collected by Zhongshan City People's Hospital Clinical Investigator Team at Zhongshan City People's Hospital, No.2 Sunwen Dong Road, Zhongshan, Guangdong 528403, P. R. China.

Investigator: Dr. Cao Yuedong, Leader; Li Qidong, Nurse A; Chen Wenqian, Nurse B.

Contact Dr. Cao Yuedong Tel: +86 760 88823818

Used Equipments DUT (Device Under Test): Transtek Wrist Blood Pressure Monitor, TMB-1117 Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-212, accuracy: ±1 mmHg and range: 0-300 mmHg.

研究设计

研究类型
Interventional
分配方式
Non Randomized
干预模型
Factorial
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • male,female,

排除标准

  • below 18 years old

结局指标

主要结局

Verify the accuracy of measure functions of device

时间窗: 10 days

Verify the accuracy of device if meet the requirements of ANSI/AAMI SP10 when it measure Systolic \& Diastolic blood pressure.

次要结局

未报告次要终点

研究者

发起方
Leo Wang
申办方类型
Other
责任方
Sponsor Investigator
主要研究者

Leo Wang

Team leader