PV Loops With CNT in Multiple Pacing Sites

Not Applicable
Not yet recruiting
Conditions
A-V Block
Bradycardia
Interventions
Device: Moderato
Registration Number
NCT05345925
Lead Sponsor
BackBeat Medical Inc
Brief Summary

In a multicenter non-randomized acute setting, eligible subjects requiring implant or replacement of an IPG (pacemaker, ICD, CRT-P, CRT-D) will be instrumented to study the effect of CNT from different locations in the RV. CNT effects on BP will be evaluated with a PV conductance catheter in the LV, and with optional BP catheters in the aorta and/or RV. CNT will be delivered by externally by the BackBeat Moderato System IPG.

Abbreviations: CNT, Cardiac Neuromodulation Therapy; IPG, Implanted Pulse Generator; ICD, Implanted Cardiac Defribillator; CRT-P or D, Cardiac Resynchronisation Therapy-Pacemaker or Defibrillator; RV, Right Ventricle; LV, Left Ventricle; BP, Blood Pressure

Detailed Description

This will be a multi-center, pilot, open-label, non-randomized acute study. The Moderato System is approved for marketing in Europe (has a CE mark) for standard pacing and the treatment of high blood pressure in patients requiring standard cardiac pacing. The therapy delivered by the Moderato® System is referred to as CNT (Cardiac Neuromodulation Therapy).

This investigation will evaluate the effect of different pacing lead positions on cardiac function and the modulation of the autonomic nervous system using Cardiac Neuromodulation Therapy (CNT) pacing signals. The Study includes both patient who have hypertension and patients that do not have hypertension. Moderato®system is currently not indicated for patients that do not have hypertension.

Subjects indicated for an implant or replacement of a device capable of pacing (pacemaker, ICD, CRT-P, CRT-D) who meet the study inclusion and exclusion criteria may participate in the study and will undergo the following procedure during their device implantation or replacement:

A Right Atrial (RA) lead will be positioned according to standard procedure of the initially planned device implant, or kept in place (in device replacement), or only temporarily introduced and used for the trial procedure. Existing ventricular leads (if any) can be used for the trial procedure, too. The ventricular lead will be temporarily positioned in the different pacing sites according to the following order: RV septal, LBBA, LBB. Pacing and CNT will be performed in each of these sites unless the site is unreachable or pacing is not possible. When trial procedures are finished, the ventricular lead will be positioned according to best practices of the implant or replacement procedure the patient is undergoing. Any lead used temporarily would be removed.

The leads will be connected to a BackBeat Moderato® System IPG, externally to the patient, using a single use, sterile Pacing System Analyzer (PSA) cable provided by the hospital. The Moderato®IPG - which is approved for chronic CNT delivery in hypertensive patients - will function in this trial as a temporary external CNT signal generator for acute application of the therapy; it will not be in direct contact with the subject.

A standard conductance catheter (CD Leycom, The Netherlands) will be introduced through the femoral artery and advanced into the left ventricle to measure cardiac volumes and pressure.

Arterial blood pressure will be measured by introducing a pressure sensor (e.g., Millar catheter) either from another femoral artery sheath or from a radial arterial line.

Right ventricular blood pressure measurements will be taken by introducing a pressure sensor (e.g., Millar catheter) to the right ventricle through an axillary or subclavian sheath.

The Moderato®IPG will be used to generate standard pacing and Cardiac Neuromodulation Therapy (CNT) signals and deliver them to the patient through the cable and leads. A range of CNT signal parameters will be used to assess the effect on sympathetic activity at the different lead positions. Ventricular pressure and volume readings from the conductance catheter, arterial pressure readings and right ventricle pressure readings will be recorded and analyzed to assess the effect of CNT signals on cardiac function, sympathetic activity and blood pressure. The effects of CNT signal over a range of parameter settings will be studied in the different lead positions.

Simultaneous recording of the ventricular and arterial pressures, ventricular volume measurements and ECG signals will be performed in every pacing site during both pacing and CNT using the LEYCOM® Inca (IntraCardiac Analyzer) system, an integrated, stand-alone system for constant monitoring of intra-ventricular pressure and volume (CE marked).

In this study, none of the Moderato® system components will be in contact with the patient. The Moderato® IPG will be used solely as an external CNT signal generator and deliver signals to the patient through the PSA cable and the pacemaker leads.

When CNT signal activation is complete, data recording will be stopped, the conductance catheter will be removed, the cable connecting the IPG to the leads disconnected, and device implantation (or replacement) will proceed normally. Any pacing lead used temporarily would be removed if not required for ongoing patient treatment.

Patients will be considered as having completed the study at the conclusion of the acute procedure unless a safety event is observed during the acute procedure in which case the patients will be followed until the event has been resolved.

Recruitment & Eligibility

Status
NOT_YET_RECRUITING
Sex
All
Target Recruitment
30
Inclusion Criteria
  • Subject is ≥ 18 years of age.
  • Subject is indicated for an implant or a replacement with a planned upgrade of a device capable of pacing (pacemaker, ICD, CRT-P, CRT-D).
  • Subject is willing and able to comply with the study procedures.
Exclusion Criteria
  • Subject is dependent on 100% ventricular pacing.
  • Subject has symptoms of heart failure, NYHA Class III or greater
  • Subject has an ejection fraction of 25% or less
  • Subject's systolic blood pressure is less than 100 mmHg on the day of implant
  • Subject has decompensated heart failure
  • Subject has significant (>2+) valvular regurgitation or any valvular stenosis.
  • Subject has permanent atrial fibrillation
  • Subject has atrial fibrillation on the day of the study.
  • Subject has hypertrophic cardiomyopathy, restrictive cardiomyopathy, or interventricular septal thickness ≥15 mm
  • Subject is on dialysis
  • Subject has a history of prior neurological events (stroke or TIA) within the past year or a neurological event (stroke) at any prior time that has resulted in residual neurologic deficit.
  • Subject has a history of autonomic dysfunction
  • Patient cannot receive heparin for any reason (such as a history of Heparin induced thrombocytopenia (HIT))
  • Women who are pregnant or breast-feeding
  • Subject cannot or is unwilling to provide informed consent.

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Experimental groupModeratoEligible subjects instrumented for CNT testing
Primary Outcome Measures
NameTimeMethod
Average changes in Systolic Blood Pressure Changethrough study completion (18 months)

When compared to standard pacing, CNT elicits a reduction in Systolic Blood Pressure (mmHg)

Secondary Outcome Measures
NameTimeMethod
Average Cardiac Volumes (End Diastolic and End Systolic volumes, stroke volumes)through study completion (18 months)

When compared to standard pacing, CNT varies cardiac volumes (ml)

Average Ejection Fraction (the percent of blood volume ejected per beat)through study completion (18 months)

When compared to standard pacing, CNT varies Ejection Fraction (EF, %)

Average Cardiac Pressures (End Systolic and Diastolic Pressures)through study completion (18 months)

When compared to standard pacing, CNT varies cardiac pressures (mmHg)

Average cardiac contractilitythrough study completion (18 months)

When compared to standard pacing, CNT varies cardiac contractility (mmHg/sec)

Average Systemic Vascular Resistancethrough study completion (18 months)

When compared to standard pacing, CNT Systemic Resistance (ml/mmHg/sec)