A Novel Ureteric Stent in Kidney Stone Patients and Oncology Patients Compared to a Conventional JJ Stent
- Conditions
- Catheter BlockageOncologyKidney Stone
- Registration Number
- NCT06815120
- Lead Sponsor
- University of Southampton
- Brief Summary
Urological stents and catheters often lead to inflammation, causing pain and infection in the urinary tract. Moreover, 80% of stents are associated with pain, negatively impacting on QoL and mental health. Offering novel designs with significantly lower E\&B leads to a reduction in UTIs and improves QoL. Reducing hospital admissions (from 3 to 1 per patient, annually) would free \>100,000 bed-nights, allowing the elderly to regain independence. Our proposed research could have a significant impact towards fulfilling the 'healthy-ageing' Grand Challenge. Additionally, the novel stent reduces prevalence of infections and therefore, of antibiotic prescriptions contributing to the Global AMR challenge.
- Detailed Description
Ureteric stents and catheters are deployed clinically as temporary measures to restore urinary drainage, in patients with kidney stones, tumours and strictures. Prevalence of these increases with age. Device-associated encrustation-and-biofilm-formation (E\&B) are key complications, leading to urinary-tract-infections (UTIs) in \>90% of stents, causing stent blockage and favouring development of antibiotic resistance.
UTIs are a significant cause of morbidity, especially among the elderly population, with 4,835 deaths in England and Wales reported in 2012. The National Institute for Health Research reported that 1-3% of all primary care consultations concern UTI-related symptoms, leading to 13.7% of antibiotic prescriptions globally. The report also revealed that \>92m people globally are diagnosed with UTIs annually.
Urological-device-associated infections significantly compromise patient' quality of life and the effectiveness of services, imposing a £2.5b annual burden on the NHS. Patients' exposure to hospital environment and general anaesthetic use impact on Quality-of-Life (QoL), resulting in unnecessary hospital bed-nights. We have developed a novel urological stent associated with significantly reduced particle deposition , potentially extending the stent's lifetime, resulting in reduced hospitalisation and improved QoL.
Urological stents and catheters often lead to inflammation, causing pain and infection in the urinary tract. Moreover, 80% of stents are associated with pain, negatively impacting on QoL and mental health. Offering novel designs with significantly lower E\&B leads to a reduction in UTIs and improves QoL. Reducing hospital admissions (from 3 to 1 per patient, annually) would free \>100,000 bed-nights, allowing the elderly to regain independence. Our proposed research could have a significant impact towards fulfilling the 'healthy-ageing' Grand Challenge. Additionally, the novel stent reduces prevalence of infections and therefore, of antibiotic prescriptions contributing to the Global AMR challenge.
Considering only onco-urological patients, \>30,000 stents are inserted every year across 200 NHS units. Due to stent failures, each patient undergoes 3 to 6 replacements, resulting in \>90,000 stent replacements. Under the current tariff-based system, hospitals are paid for each intervention, costing the NHS \>£3,500, bringing the total cost of replacements to \~£315m annually. Unnecessary replacements increase the number of bed-nights and use of anaesthetics. A longer-lasting stent means that each patient comes to the hospital ideally only once, reducing their exposure to the hospital environment and anaesthetic-associated risks. Additionally, under the current model hospitals would improve their quality of service and save at least 2 bed-nights per patient. The NHS is moving towards a "block contract" model, meaning that hospitals will receive a lump sum for the year and they'll regularly monitor budgets and improve cost efficiency where possible. Improved stents will allow the same urology budget to deliver more healthcare, reducing theatre time for stent insertion-and-replacement. Our innovative stent design offers a potential annual saving of \>£210m for the NHS, allowing hospitals to reallocate their resources. Today E\&B is still a major determinant of stent failure and associated side effects. We aim to address this challenge through our novel stent, and if equivalent safety and improved efficacy are demonstrated, anticipate market launch via third-party suppliers 2-3 years post project.
Solutions to break this pathway (stent presence \>\> inflammation, pain and infection) have been developed, including materials (e.g. metallic alloys, polymers, biodegradable and drug-eluting materials), coatings (e.g. heparin, chitosan, hydrogel, carbon) and shapes (e.g. double-J, loop, mesh, string, expandable) to improve the efficacy and safety of stents. Despite all these advances, there is still a significant prevalence of E\&B. This is also due to the lack of studies correlating fluid dynamic metrics with deposition of particles causing E\&B. The lead applicant's PhD investigated the mechanisms of particle deposition in urological stents, and successfully determined fluid dynamic parameters governing particle deposition on ureteric stents and catheters. His research led to developing a stent with specially shaped side-holes that prevent stagnation points (i.e., areas of low flow that cause particles to settle and E\&B). These developments demonstrated \>80% reduction in particle deposition at side-holes, in-vitro. Their innovative architecture can be implemented on stents and catheters. The technology, including the stent design and manufacturing process, is protectable - Patent# WO2019048860A1, WIPO(PCT). These results were further validated in an animal study (6 pigs: 3 novel-design-stent vs 3 conventional-design-stent for a 4-weeks period) at Stone-Centre at Vancouver-General-Hospital (VGH), a centre of excellence in animal studies on urological products. The study (available upon request) concluded that the functionality of stents with novel side-holes is the same as that of standard stents in terms of safety. Specifically, the novel stent does not result in increased risk of irritation, inflammation and hydronephrosis. Furthermore, SEM/EDX imaging showed that the novel design decreased the build-up of particles on the stent surface. Thus, it demonstrated significant potential for the new side-hole configuration to change patterns of particle deposition on the stent's surface, decreasing encrustation
Recruitment & Eligibility
- Status
- NOT_YET_RECRUITING
- Sex
- All
- Target Recruitment
- 50
- Aged 18 years or over
- Ureteric stents clinically indicated either due to kidney stones or abdominal/pelvic cancers compressing ureters
- Previous experience with ureteric stents
- Awaiting insersion/replacement of stents
- Ability to give consent
- Ability to interact with study documentation
- Sufficient English to complete study documentations and questionnaires
- Expected survival <4months
- Unfit for stent insertion
- Unable to comply with study processes Pregnancy
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Primary Outcome Measures
Name Time Method To determine rates of stent failure From enrolment to stent removal (4 weeks for Kidney cohort), (25 weeks for Oncology cohort) Outcome measure: Stent failure, defined as any of: a): a stent change earlier than planned (oncology only); b) need for additional surgical or radiological intervention; c): kidney failure (evidenced by acute kidney injury (AKI) on blood tests (eGFR or creatinine kidney function tests) and/or worsening hydronephrosis on imaging
Summary method: Frequency and percentage of people experiencing any stent failure
- Secondary Outcome Measures
Name Time Method To assess extent of encrustation and biofilm (E&B) From enrolment to stent removal (4 weeks for Kidney cohort), (25 weeks for Oncology cohort) Outcome measure:
* Cell culturing (CFU counts per ml per bacterial microorganism)
* Flow study (pressure value): A section of the retrieved stent will be used for flow study. The part will be placed into a larger diameter tube. DI water/Saline is injected using a pump into the inlet of the external tube and pressure sensor is connected to the inlet tube. Pressure readings will be captured per stent.
* Dwell time
Summary method:
* Imaging (selected samples of stents will be examined using non-contact, non-destructive SEM (scanning-electron-microscopy) and/or episcopic-differential-interference-contrast (EDIC) microscopy allowing imaging of the surface with the biofilm in-situ and colonisation by bacteria - when needed)
* Mean and standard deviation or median and quartiles of pressure readingsTo determine whether the novel stent leads to better clinical outcomes From enrolment to completion of patient questionnaires 2-3 weeks post surgery 2 (4 weeks for Kidney cohort), 2-3 weeks post surgery 2 (~25 weeks for Oncology cohort) Outcome measure:
* Individual component of the primary outcome
* Urinary symptoms
* Infections
Summary method:
* As per primary outcome
* Frequency and percentage of participants experiencing each symptom
* Frequency and percentage of people with antibiotic
* use, readmission, urosepsis, ITU, death; number of infections per unit of time stent is in useTo assess impact of the novel stent on quality of life Within first month of stent removal for participants (4 weeks for Kidney cohort), (25 weeks for Oncology cohort) Outcome measure:
Questionnaires at:
* Baseline
* While stent is in situ
* Post-surgery
Summary method:
Appropriate descriptive summaries (e.g., median and quartiles) per timepointTo understand patient experience in having the novel ureteric stent inserted and reason for participation in the trial From enrolment to stent removal (4 weeks for Kidney cohort), (25 weeks for Oncology cohort) Outcome measure:
30-45-minute Qualitative interviews with Oncology and Kidney stone cohorts
Summary method:
An inductive thematic analysis of interview transcripts