Assessment of Right Ventricular Volume Using the Ventripoint Medical System in Patients With Tetralogy of Fallott Following Repair; A Comparison Study to cMRI
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 109
- 试验地点
- 6
- 主要终点
- The primary endpoint of this study is the percent difference between the VentriPoint Medical System (VMS) and cMRI for estimating the end diastolic and end systolic right ventricular volumes (EDV and ESV) in subjects following repair for TOF.
研究概览
简要总结
Assessment of Right Ventricular Volume using the VentriPoint Medical system in patients with Tetralogy of Fallot following repair; a comparison study to cMRI.
The objective of this study is: To evaluate the accuracy of the VentriPoint Medical System to calculate right ventricular volumes in patients with Tetralogy of Fallot following surgical repair.
Secondary objectives are: To validate Inter and Intra observer variability at 3 clinical sites.
The analyses will be the same for EDV and ESV. For either right ventricular volume, the primary effectiveness measure is the % difference between VMS and cMRI results, i.e. (VMS-cMRI)*100%/average of VMS and cMRI results. There will be two null hypotheses for the primary analysis:
H0+: true mean % difference > 10% and H0-: true mean % difference < -10%
The observed mean % difference will be presented with 95% confidence intervals. The VMS right ventricular volume estimates will be regarded as equivalent to cMRI estimates if both H0+ and H0- are rejected at a 1-sided 0.025 level by a paired t test for both EDV and ESV. Since H0+ and H0- cannot both be true, the total type I error rate for each measure is 0.025 and the overall type I error rate for both EDV and ESV is 0.05.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Prospective
入排标准
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patients with Tetralogy of Fallot following surgical repair
- •Patients who can be expected to lie motionless during imaging
排除标准
- •Lack of informed consent
- •Surgical repair for Tetralogy of Fallot with RV-PA conduit
- •Known arrhythmia that interferes with image acquisition.
- •Implanted cardiac defibrillator, pacemaker, or other devices containing ferromagnetic materials
- •Pregnant woman
- •Contraindications for MRI
结局指标
主要结局
The primary endpoint of this study is the percent difference between the VentriPoint Medical System (VMS) and cMRI for estimating the end diastolic and end systolic right ventricular volumes (EDV and ESV) in subjects following repair for TOF.
时间窗: 4 months
The trial will be regarded as successful if it demonstrates the mean VMS-cMRI percent difference to be \<10% and \>-10% at a 1-sided 0.025 statistical significance level for each of EDV and ESV, with no safety concerns for the VMS procedure.