TENS in Optimizer(R) Patients

Phase 4
Completed
Conditions
Heart Failure
Interventions
Device: Stimulette r4x
Registration Number
NCT02692690
Lead Sponsor
Medical University of Vienna
Brief Summary

Systemic skeletal muscle atrophy (cachexia) is a typical side effect of chronic heart failure.

Patients with cachexia can benefit from transcutaneous electrical neuromuscular stimulation (TENS). It is applied to strengthen the musculature and improve exercise tolerance and quality of life.

It is unclear if application of TENS can cause electromagnetic interference (EMI) in patients with Optimizer® pacemaker systems for improvement of heart failure.

This pilot safety study enrolled 6 patients with chronic heart failure and reduced left ventricular ejection fraction. All patients had an Optimizer® pacemaker system implanted for a minimum of 6 months, no patients had any ventricular arrhythmic episode 3 months prior to the study, and all were in clinically stable condition and willing to participate in this study.

Neuromuscular stimulation was administered via TENS in 4 different modes (HF, LF, Burst, MF) on the right and left musculus trapezius and on both musculi quadriceps femoris. All patients were tested for EMI without or with cardiac contractility modulation (CCM) delivery of the Optimizer®-system respectively.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
Male
Target Recruitment
6
Inclusion Criteria
  • >18 years, with implantation of the Optimizer ® >6 months ago, clinically stable condition and without any ventricular arrhythmic episode in the last 3 months.
Read More
Exclusion Criteria
  • Local swelling or edema at the site of stimulation, intolerance to electrical stimulation, or presence of other metallic implants at the site of stimulation
Read More

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
before-afterStimulette r4xTENS stimulation neck and thigh
Primary Outcome Measures
NameTimeMethod
The patients with CCM-system are assessed for EMI during TENS-application6 months

Electromagnetic interference can cause artificial sensing in pacemaker or ICD system. These artefacts may have the effect of suppressing the pacemaker pulses sensing of tachycardia in ICD.

The patients will be exposed to electrical stimulations, which will be of different forces on the neck and the upper thigh with r4x. These impulses and the effects will be noted down. It will be investigated if there is an interference with the optimizer ® depending on the different sites of the electrical stimulations and the different impulse forces.

Many patients with an Optimizer ® have simultaneously a pacemaker - or ICD implant. A special extra monitor will note if there are probable interferences with these pacemaker- or ICD implants.