Study C3S : Comparison Among Healthy Subject
- Conditions
- Healthy Volunteers
- Registration Number
- NCT03614624
- Lead Sponsor
- University Hospital, Clermont-Ferrand
- Brief Summary
The aim of the C3S study is research an equivalence with longitudinal, radial and circumferential, global and segmental strain value, in left Ventricular in 2 and 3 dimension between two echograph General Electric Vivid 9 (GE Vingmed, Horton, Norway) et Philips Epiq 7.
- Detailed Description
Healthy volunteer will be inform about the study by email and press. They could join the clinical research team by email and phone to ask more questions about the study. Then the investigator, verify the eligibiliy of the volunteers and schedule an appointment for the visit of inclusion. During this visit, the patient's consent will be collected. One echocardiography will be complete on echograph General Electric Vivid 9 to collect the clinical data necessary for the study and to eliminate the presence of criteria of exclusions, the acquisition of the loops of ultrasound will be realized on both echographs by a first investigator. A second investigator will make the post-treatment of the ultrasound images on consoles ECHOPAC PC version 201, General Electric Medical System and Q Station version 3.5, Philips Healthcare in two different times, to obtain the values of the various strain.
Furthermore, an clinical exam, an ECG, a questionnary will be realized.
Recruitment & Eligibility
- Status
- UNKNOWN
- Sex
- All
- Target Recruitment
- 66
- Patients over 18 years' old
- Without any cardiac pathology
- Not taking long-term treatments, except oral contraception
- Consented to participate in the study
- Existence of heart disease : myocardial revascularization, primitive cardiopathy (dilated or hypertrophic cardiopathy), moderate or severe valvulopathy, pulmonary hypertension, congenital heart disease operated or not, previous cardiac heart failure, previous rhythm disturbance and/or pacemaker, hypertension treated or untreated, peripheral artery disease : occlusive arteriopathy of the lower limbs, peripheral arterial aneurysm, stenosis or peripheral arterial occlusion.
- Practicing intensive sport (more than 6 hours per week)
- Untreated type 2 or 1 diabetes , treated dyslipidemia, chronic alcoholic intoxication (men more than 3 units of alcohol per day and women more than 2 units per day)
- Existence of heart disease or previous heart disease
- Pregnant or lactating women,
- Dependent adult
- Patients minor,
- Decline the study,
- Existence of a psychiatric pathology
- Inability to grant its informed consent
- No affiliation to a social security scheme
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Primary Outcome Measures
Name Time Method Value of the longitudinal strain value, in left Ventricular in 2 and 3 dimension. 1 year The acquisition of the ultrasound loops on both echographs is realized at the same time. The various values of strain are obtained after post-treatment of the ultrasound loops on ECHOPAC PC version 201 software for the echograph General Electric and the Q Station version 3.5 software for the echograph Phillips. The post-treatment on both software are made in two different times to avoid possible biases.
Values of the radial strain value, in left Ventricular in 2 and 3 dimension. 1 year The acquisition of the ultrasound loops on both echographs is realized at the same time. The various values of strain are obtained after post-treatment of the ultrasound loops on ECHOPAC PC version 201 software for the echograph General Electric and the Q Station version 3.5 software for the echograph Phillips. The post-treatment on both software are made in two different times to avoid possible biases.
Values of the circumferential strain value, in left Ventricular in 2 and 3 dimension. 1 year The acquisition of the ultrasound loops on both echographs is realized at the same time. The various values of strain are obtained after post-treatment of the ultrasound loops on ECHOPAC PC version 201 software for the echograph General Electric and the Q Station version 3.5 software for the echograph Phillips. The post-treatment on both software are made in two different times to avoid possible biases.
Values of the global strain value, in left Ventricular in 2 and 3 1 year The acquisition of the ultrasound loops on both echographs is realized at the same time. The various values of strain are obtained after post-treatment of the ultrasound loops on ECHOPAC PC version 201 software for the echograph General Electric and the Q Station version 3.5 software for the echograph Phillips. The post-treatment on both software are made in two different times to avoid possible biases.
Values of the segmental strain value, in left Ventricular in 2 and 3 1 year The acquisition of the ultrasound loops on both echographs is realized at the same time. The various values of strain are obtained after post-treatment of the ultrasound loops on ECHOPAC PC version 201 software for the echograph General Electric and the Q Station version 3.5 software for the echograph Phillips. The post-treatment on both software are made in two different times to avoid possible biases.
- Secondary Outcome Measures
Name Time Method Values of segmental 2D of the right ventricle between two echographs (General Electric and Philips). 1 year The acquisition of the ultrasound loops on both echographs are realized at the same time. The various values of strain are obtained after post-treatment of the ultrasound loops on ECHOPAC PC version 201 software for the echograph General Electric and the Q Station version 3.5 software for the echograph Phillips. The post-treatment on both software are made in two different times to avoid possible biases.
Values of strain 2D of the left atrium between two echographs (General Electric and Philips). 1 year The acquisition of the ultrasound loops on both echographs are realized at the same time. The various values of strain are obtained after post-treatment of the ultrasound loops on ECHOPAC PC version 201 software for the echograph General Electric and the Q Station version 3.5 software for the echograph Phillips. The post-treatment on both software are made in two different times to avoid possible biases.
Values of longitudinal global strain of the right ventricle between two echographs (General Electric and Philips). 1 year The acquisition of the ultrasound loops on both echographs are realized at the same time. The various values of strain are obtained after post-treatment of the ultrasound loops on ECHOPAC PC version 201 software for the echograph General Electric and the Q Station version 3.5 software for the echograph Phillips. The post-treatment on both software are made in two different times to avoid possible biases.
Trial Locations
- Locations (1)
CHU de Clermont-Ferrand
🇫🇷Clermont-Ferrand, Auvergne, France